Under the gait mode, user can choose among smart mode, normal mode, and manual mode. Click the
“smart mode” position to pop up the mode selection list and select one.
a. Smart mode: The host automatically calculates the tilt angle A of the electrical stimulation and the tilt
angle B of shutting down electrical stimulation according to the gait data of the first four steps of the
patient. The parameters that can be adjusted are the electrical stimulation intensity, frequency and pulse
width in the lower level interface of the “Parameter Settings”.
b. Normal mode: The host performs electrical stimulation according to the set parameters. The parameters
that can be adjusted are the electrical stimulation intensity, and the frequency, pulse width, tilt angle A, tilt
angle B, duration, delay time, ramp up and ramp down in the lower-level interface of the “Parameter
Setting”.
c. Manual mode: After starting this mode, the host performs electrical stimulation according to the
electrical stimulation button of the interface; press the button to start electrical stimulation, stop the
electrical stimulation by the release, and continue to hold the electrical stimulation for continuous output.
The parameters that can be adjusted are the electrical stimulation intensity and the frequency and pulse
width in the lower level interface of the “Parameter Settings”.
Training mode
The training mode performs electrical stimulation based on a combination of parameters of the selected
training mode. The parameters combine 9 fixed combinations and one custom. The electrical stimulation
can be adjusted after the fixed mode starts, and other parameters cannot be adjusted. All parameters can
be adjusted under custom.
Start: The App sends the training mode parameters and mode start command to the host, receives the host
reply, and enters the training work interface.
Pause: The App sends a training mode pause (stop) command to the host, receives a host reply, and the
training is suspended.
Continue: The App sends the training mode (start), at which point the current intensity given in the
parameter settings is the current strength before the pause. After receiving the host response, the training
work continues.
Stop: After long press and stop for 2s, send the training mode stop command to the host, receive the host
reply, and the training mode stops.
1-9 Built-in 9 training modes with fixed parameters; custom parameters can be adjusted.
Evaluation mode
Evaluation mode: Collect the angle data generated by the host during the patient's walking. The collection
duration is 60s by default and the maximum is 90s. The acquisition begins to draw an angle waveform
and displays the countdown of the acquisition time. Can stop in the middle. The acquisition is completed
to display the evaluation results.
Start: The App sends an evaluation mode to start the command to the host, receives the host reply, and
enters the evaluation work interface.