User's Manual

20
Q2. After turning on the training mode or gait mode on the unit, although the indicator light is on, there is no electrical
stimulation reaction.
A. -Check if the host is tied and close to the skin.
-Check if the current intensity set by the host is adjusted to the appropriate intensity.
-Add some water to the surface of the electrode sheet to increase the conductivity of the electrode sheet.
Q3. What should I do if the skin in the area covered by the electrode and the host band is severely red, stinging or
allergic?
A. Stop using it immediately. After observing for a period of time, if no abnormality is found, wait until the skin is
completely improved before continuing to use the device. Remember to regularly ventilate the skin covered by the host.
Q4. The unit display battery icon is turned off after 1 second interval flashing.
A. This indicates that the host is running low and needs to be recharged. It takes about 3 hours for the unit to charge;
after the battery is fully charged, the unit can last for about 10 hours. When the battery is low, please charge it in time.
Q5. The host LED light flashes quickly, what should I do if the main unit display “ ”, “TBD Icon” flashes
alternately?
A. This indicator is a reminder that the electrode is off. Please check if the host is fastened. Or the skin is too dry and
needs to moisturize the skin. After re-fixing to the leg, press the main power stimulation adjustment button or press the
stimulation adjustment button in APP software to continue.
Q6. What happens when there is sporadic strong electrical stimulation?
A. -The surface of the electrode is not wet enough to sprinkle some water on the surface.
-Check if the skin in the area covered by the electrode is red or has a wound.
-Check if the cuff of the main unit is loose and the position of the electrode is accurate.
Q7. Why is there no electrical stimulation at the time of electrical stimulation?
A. Usually because the cuff position has been moved or the gait mode has changed. Please wear the cuff again or reset
the walking parameters.
Q8. Can I use oil or lotion on my legs?
A. No, please make sure the skin is clean before using the stimulator and fully wet the adhesive surface of integrated
electrode.
8. Product Configuration
8.1 Product specifications
Communication method: Bluetooth 4.0
Communication frequency: 2.4-2.4835GHz
8.2 Stimulator
Power Supply 3.7V rechargeable lithium battery or DC 5V, power adapter
Classification Type BF applied part, internally powered equipment
Shutdown current ≤10μA
Working current ≤120mA