PM-9000 Express Patient Monitor Service Manual
Service Manual (V 1.0) Copyright Statement SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns all rights to this unpublished work and intends to maintain this work as confidential. Mindray may also seek to maintain this work as an unpublished copyright. This publication is to be used solely for the purposes of reference, operation, maintenance, or repair of Mindray equipment. No part of this can be disseminated for other purposes.
Service Manual (V 1.0) adjusting the volume to very low or completely muting the sound may result in the disaster to the patient. The most reliable way of monitoring the patient is at the same time of using monitoring equipment correctly, manual monitoring should be carried out. This multi-parameter patient monitor is intended for use only by medical professionals in health care institutions. To avoid electrical shock, you shall not open any cover by yourself.
Service Manual (V 1.0) Warranty THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING WARRANTIES OF MERCHANT ABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE.
Service Manual (V 1.0) Return Policy Return Procedure In the event that it becomes necessary to return a unit to Mindray, the following procedure should be followed: 1. Obtain return authorization. Contact the Mindray Service Department and obtain a Customer Service Authorization (Mindray) number. The Mindray number must appear on the outside of the shipping container. Return shipments will not be accepted if the Mindray number is not clearly visible.
Service Manual (V 1.0) Safety Precautions 1. Meaning of Signal Words In this manual, the signal words DANGER, WARNING, and CAUTION are used regarding safety and other important instructions. The signal words and their meanings are defined as follows. Please understand their meanings clearly before reading this manual. Signal word Indicates an imminently hazardous situation which, if not DANGER avoided, will result in death or serious injury.
Service Manual (V 1.0) current at power ON, the circuit breakers and fuses of the building’s supply circuit may be tripped. CAUTION: 1. Malfunctions due to radio waves (1) Use of radio-wave-emitting devices in the proximity of this kind of medical electronic system may interfere with its operation. Do not bring or use devices which generate radio waves, such as cellular telephones, transceivers, and radio controlled toys, in the room where the system is installed.
Service Manual (V 1.0) Symbols Be Careful Protective earth ground Indicates that the instrument is IEC-60601-1 Type CF equipment. The unit displaying this symbol contains an F-Type isolated (floating) patient applied part providing a high degree of protection against shock, and is suitable for use during defibrillation.
Service Manual (V 1.0) Contents CHAPTER 1 ABOUT THE PRODUCT ...........................................................................................1 1.1 INTRODUCTION ........................................................................................................................1 1.2 APPLICATION ............................................................................................................................1 1.3 ENVIRONMENT ........................................................
Service Manual (V 1.0) CHAPTER 6 MAINTENANCE AND CLEANING ..........................................................................66 6.1 MAINTENANCE .......................................................................................................................66 6.2 CLEANING ..............................................................................................................................66 6.3 CLEANING REAGENT ......................................................................
Chapter 1 Chapter 1 1.1 About the Product About the Product Introduction The PM-9000 Express Patient Monitor, a portable and accessible patient monitor, is supplied by rechargeable batteries or external AC power, which applies to adults, pediatric and neonates. You can select different configurations as required. Besides, the PM-9000 Express can be connected with the central monitoring system whereby a monitoring network will be formed.
Chapter 1 About the Product processes them and displays them on the screen. You can set the alarm limit as required. When the monitored parameter exceeds the preset alarm limit, the patient monitor will start the alarm function. In addition, you can control the patient monitor through the control panel. Usually, patient monitors are seen in some clinical areas in hospital, such as ICU, CCU, intensive care units for heart disease patients, operating rooms, emergency departments and observation wards.
Chapter 1 About the Product ISO (Isoflurane) ENF (Enflurane) SEV (Sevoflurane) DES (desflurane) Airway Respiration Rate (rpm: Respiration Per Minute): AwRR Minimum Alveolar Concentration (MAC) 4 AG waveforms (CO2, N2O, O2, AA) The PM-9000 Express provides the functions of audio/visual alarm, trend graphic storage and output, NIBP measurement, alarm event identification, large font screen, defibrillator synchronization, oxyCRG recall, drug calculation, etc. 1.3 Environment 1.3.
Chapter 2 Principles Chapter 2 Principles 2.1 General The intended use of the PM-9000 patient monitor is to monitor a fixed set of parameters including ECG, RESP, SpO2, NIBP, TEMP, IBP, CO and CO2 (IBP, CO and CO2 are optional). It consists of the following functional parts: Parameter measurement; Main control part; Man-machine interface; Power supply; Other auxiliary functions; These functional units are respectively detailed below. Figure 2-1 Structure of the PM-9000 Express 2.1.
Chapter 2 Principles parameter measurement part of the PM-9000 Express patient monitor consists of the measurement probe, parameter input socket assembly, NIBP assembly and the main control board.
Chapter 2 Principles Figure 2-2 Functional structure of the PM-9000 Express 6
Chapter 2 Principles The PM-9000 Express PCB connection is shown in the following figure. Figure 2-3 PCB connection Basic functions and working principles of modules are described in the following sections. 2.2.1 Main Board 2.2.1.1 General The main board is the heart of the patient monitor.
Chapter 2 Principles 2.2.1.2 Principle diagram Figure 2-4 Working principle of the main board 2.2.1.3 Principle The main board is connected with external ports, including the power input port, multi-way serial port, TFT display interface, analog VGA interface, network port and analog output port. Besides, on the main board is also a BDM interface reserved for the software debugging and software downloading. CPU System CPU is the core part of the main board.
Chapter 2 Principles In addition, FPGA also extends multiple serial ports, which communicate with peripheral modules. FPGA transfers the received data to CPU through the bus; CPU delivers data to FPGA through the bus, and then the FPGA transfers those data to the peripheral modules. Watchdog When powered on, watchdog provides reset signals for CPU, FPGA and Ethernet Controller. The patient monitor provides the watchdog timer output and voltage detection functions. 2.2.2 ECG/RESP/TEMP Module 2.2.2.
Chapter 2 Principles rejects the common-mode interference signal. The low-pas filtering circuit filters the high-frequency interference signal beyond the frequency band of the ECG signal. The PACE signal refers to the ECG pace signal. It has significant interference to the ECG signal detection. The PACE rejection circuit can rejects the PACE signal, which helps the ECG signal detection.
Chapter 2 Principles 2.2.3 CO/IBP Module 2.2.3.1 General This module provides the function of measuring two parameters: Cardiac Output (CO) and Invasive Blood Pressure (IBP). 2.2.3.2 Principle diagram Figure 2-6 Working principle of the CO/IBP module 2.2.3.3 Principle This module collects the CO/IBP signal through the transducers, processes it and sends it to the main board throgh the serial port. CO Signal Process Network The CO parameter is measured with the thermal dilution method.
Chapter 2 Principles 2.2.4 SpO2 Module 2.2.4.1 General This module provides the function of measuring the Pulse Oxygen Saturation (SPO2). 2.2.4.2 Principle diagram Figure2-7 Working principle of the SpO2 module 2.2.4.3 Principle The SpO2 measurement principle 1. Collecting the light signal of the red light and infrared transmitting through the finger or toe which is pulsing; 2. Processing the collected signal to get the measured result.
Chapter 2 Principles Implementing the logical control of all the circuits; Implementing the data processing for the SpO2 parameter; Implementing the communication with the main board. Power & Signal isolate Circuit Isolating the external circuits to ensure the safety of human body; Supplying power for all circuits; Implementing the isolation communication between the CPU System and the main board. 2.2.5 NIBP Module 2.2.5.
Chapter 2 Principles Motor Drive Circuit This circuit controls the action of the air pump. It, together with the Valve Drive Circuit, implements the inflation and deflation of the cuff. Besides, it provides the status signal of the motor for the A/D conversion part. NIBP Signal Process Network The NIBP signal is the differential input signal.
Chapter 2 Principles matrix data, sends them to the heat-sensitive printer, and drives the printer. Step Motor Drive Circuit There is a step motor on the heat-sensitive printer. The step motor drives the paper. This circuit is used to drive the step motor. Printer Status Detect Circuit This circuit detects the status of the heat-sensitive printer, and sends the status information to the CPU system.
Chapter 2 Principles Controlling the status of LED; Controlling the audio process circuit; Regularly resetting the Watchdog timer; Communicating with the main board. Audio Process Circuit This circuit generates audio signals and drives the speaker. Watchdog When powered on, the Watchdog provides the reset signal for CPU. The patient monitor provides the watchdog timer output and voltage detection functions. 2.2.8 Power PCB 2.2.8.1 General This module provides DC working current for other boards. 2.2.8.
Chapter 2 Principles boards. 12V DC/DC This part converts the DC voltage to the stable 12V DC voltage and supplies it for the external boards. Power Switch Circuit This circuit controls the status of the 5V DC/DC part and the 12V DC/DC part, thus to control the switch of the patient monitor. Voltage Detect Circuit This circuit detects the output voltages of the circuits, converts the analog signal to the digital signal, and sends the digital signal to the main board for processing. 2.3 2.3.
Chapter 2 Principles 2.3.
Chapter 2 Principles 2.3.3 System Function The system tasks can be classified as follows. Figure 2-13 System task 2.4 System Parameter 2.4.
Chapter 2 Principles are transferred to the main board through the adapter board, thus to process and display the data and waveforms. Commands from the main board, as well as the status information of modules, are transferred through the adapter board. In addition, the adapter board adapts and changes the power supply. The structure of the whole system is shown in the following figure.
Chapter 2 Principles 1) Input circuit: The input circuit protects the ECG input level, and filters the ECG signals and external interference. The ECG electrode is connected to the input circuit through the cable. 2) Buffer amplifying circuit: This circuit ensures extremely high input impedance and low output resistance for ECG.
Chapter 2 Principles of the patient; 2) Manual/Auto/Continuous mode: The manual measurement is also called single measurement; in this mode, only one measurement is done after being started. In the auto measurement mode, the measurement can be done once within the selected period, with the interval being 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240 or 480 minutes.
Chapter 2 Principles algorithm. 2.4.7 CO CO measurement principle: The thermal dilution method is widely used in the clinical CO monitoring. Introduce a floating catheter into the pulmonary artery through the right atrium, and inject the physiological saline into the right atrium through the catheter whose front end is connected to the temperature transducer. When the cold liquid mixes with the blood, there will be a change of temperature.
Chapter 2 Principles gases, multiple infrared filters are required in the AG module. The oxygen does not absorb the infrared within the above-mentioned wave band. Therefore, the oxygen is measured based on its paramagnetism. Inside the transducer of the O2 module, there are two crystal balls full of nitrogen. They are suspended in the symmetrical magnetic field, and designed to point to the strongest outgoing part of the magnetic field. Outside the balls is the paramagnetic oxygen.
Chapter 3 Product Specifications Chapter 3 Product Specifications 3.1 Safety Classifications Type of protection against electric shock Class I with internal electric power supply.
Chapter 3 Product Specifications 3.
Chapter 3 Product Specifications 3.3 Power Source Specifications AC mains Input voltage 100 to 240V Frequency 50/60Hz Power 140VA Fuse T 3A Internal battery Number of batteries 2 Type Sealed lead-acid battery or lithium-ion battery Time to shutdown 5 to 15min (after the first low-power alarm) Sealed lead-acid battery Nominal voltage 12VDC Capacity 2.
Chapter 3 Product Specifications 3.4 Hardware Specifications Physical Size 318 × 270 × 137mm (width×height×depth) Different due to different configurations Weight Standard configuration: 4.7kg Maximum weight: ≤ 7.5kg Display Type Color TFT LCD Size 12.
Chapter 3 Product Specifications pressed. Button 7 buttons: POWER, MAIN, FREEZE, PAUSE, RECORD, NIBP, MENU Connectors Power supply 1 AC power connector Parameter ECG, RESP, TEMP, SPO2, NIBP, IBP, CO, CO2, AG Network 1 standard RJ45 network connector, 100 BASE-TX VGA 1 standard color VGA monitor connector, 15-PIN D-sub Auxiliary output 1 BNC connector Equipotentiality 1 equipotential grounding connector 3.5 Wireless network Standards IEEE 802.11b, Wi-Fi compatible Frequenct range 2.
Chapter 3 Product Specifications 3.6 Data Storage Trend data Long trend: 96 hours, resolution 1min, 5 min or 10 min. Short trend: 1 hour, resolution 1 s or 5 s. Alarm events 70 alarm events and associated waveforms (with user selectable waveform length 8s, 16 or 32). ARR events 80 ARR events and associated waveforms with 8s wavelength. NIBP measurements 800 NIBP groups, including systolic pressures, mean pressures, diastolic pressures and measurement time.
Chapter 3 Product Specifications 3.7 Signal Output Specifications Standards Meets the requirements of EC60601-1 for short-circuit protection and leakage current Output impedance 50Ω ECG analog output 0.05 to 100Hz(12-lead: 0.
Chapter 3 Product Specifications 3.8 ECG Specifications Lead type 3-lead (1 channel): I, II, III 5-lead (2 channels): I, II, III, aVR, aVL, aVF and V 12-lead (8 channels): I, II, III, avR, avL, avF, V1-V6 Lead naming style AHA, EURO Sensitivity selection 1.25mm/mV (×0.125), 2.5mm/mV (×0.25), 5mm/mV (×0.5), 10mm/mV (×1), 20mm/mV (×2) and auto Sweep speed 12.5mm/s, 25mm/s, 50mm/s Bandwidth (– 3dB) Common mode rejection Diagnostic mode: 0.05 to 100Hz (12-lead: 0.
Chapter 3 Product Specifications Response time to heart rate changes Meets the requirement of ANSI/AAMI EC13-2002: Section 4.1.2.1 f). Less than 11 sec for a step increase from 80 to 120 BPM Less than 11 sec for a step decrease from 80 to 40 BPM When tested in accordance with ANSI/AAMI EC13-2002 Section 4.1.2.1 g, the response time is as follows. Response time of tachycardia alarm Figure 4ah – range: 15.7 to 19.2s, average: 17.4s 4a – range: 5.7 to 8.5s, average: 7.5s 4ad – range: 3.6 to 5.
Chapter 3 Product Specifications 3.9 RESP Specifications Measurement technique Thoracic impedance Lead Optional: lead I and lead II; default lead II Differential input impedance > 2.5MΩ Respiration impedance test range 0.3 to 3Ω Excitation current < 300µA Baseline impedance range 200 to 2500Ω (using an ECG cable with 1kΩ resistance) Bandwidth 0.2 to 2Hz (-3 dB) Sweep speed 6.25 mm/s, 12.
Chapter 3 Product Specifications 3.10 SpO2 Specifications Mindray SpO2 Module SpO2 Measurement range 0 to 100% Resolution 1% 70 to 100%: Precision 70 to 100%: 70 to 100%: 0% to 69%: Refreshing rate ±2 % (adult/pediatric, non-motion conditions) ±3 % (neonate, non-motion conditions) ±3 % (in motion conditions) Undefined.
Chapter 3 Product Specifications PR Measurement range 25 to 240bpm Resolution 1bpm Precision Refreshing rate ±3bpm (non-motion conditions) ±5bpm (in motion conditions) 1s Nellcor SpO2 Specifications SpO2 measurement range and precision Sensor Range Precision* MAX-A, MAX-AL, MAX-N, MAX-P, MAX-I and MAX-FAST 70 to 100% ±2% 0% to 69% Undefined OxiCliq A, OxiCliq N, OxiCliq P, OxiCliq I 70 to 100% ±2.
Chapter 3 Product Specifications 3.
Chapter 3 Product Specifications 3.12 TEMP Specifications Number of channels 2 Displayed parameters T1, T2 and TD Measurement range 0 to 50°C (32 to 122°F) Resolution 0.1°C Precision Update period Minimum time for accurate measurement 0.1°C (excluding the sensor) ±0.
Chapter 3 Product Specifications 3.
Chapter 3 Product Specifications 3.14 CO Specifications Measurement technique Thermal dilution Calculated parameter CO, hemodynamics Measurement range Resolution Precision Alarm range CO 0.1 to 20l/min TB 23 to 43°C TI 0 to 27°C CO: 0.1 l /min TB, TI: 0.1°C CO: ±5% or ± 0.1 l /min 0.
Chapter 3 Product Specifications 3.15 CO2 Specifications Measurement technique Infrared absorption technique Measurement mode Sidestream, microstream or mainstream (optional) Displayed parameter EtCO2, FiCO2, Respiration Rate CO2 function Meet the requirements of EN864 and ISO9918.
Chapter 3 Product Specifications * Conditions for measurements in typical precision: The measurement is started after the preheating mode of the module; Ambient pressure: 750mmHg to 760mmHg; room temperature: 22℃ to 28℃; The gas under test is dry, and the balance gas is N2; The deflation rate is 150ml/min, the respiration rate is no greater than 30BrPM, with a fluctuation less than ±3BrPM, and the inhale interval/exhale interval is 1:2; In other conditions, the measurement precision should meet the require
Chapter 3 Product Specifications Welch Allyn CO2 Specifications CO2 measurement range Precision* 0 to 99mmHg 0 to 40 mmHg: ±2mmHg 41 to 76 mmHg: ±5% 77 to 99 mmHg: ±10% Resolution 1mmHg Refreshing rate 1s Start-up time < 80s (ambient temperature: 25℃; preheating power of transducer: 5W) Response time 100ms (10% to 90 %) Calibration Daily calibration is unnecessary Calibration stability There is a difference (< 1%) from the precison criteria after a 12-month continuous service time Alar
Chapter 3 Product Specifications 3.
Chapter 3 Product Specifications Sev Enf, Iso, Hal O2 (Optional) AwRR Alarm range Apnea alarm delay 5 to 10 ±0.4 10 to 15 ±0.6 15 to 18 ±1 >18 Not specified 0 to 1 ±0.15 1 to 5 ±0.2 5 to 8 ±0.4 >8 Not specified 0 to 1 ±0.15 1 to 5 ±0.2 >5 Not specified 0 to 25 ±1 25 to 80 ±2 80 to 100 ±3 2 to 60 BrPM ±1 BrPM > 60 BrPM Not specified 0 to 10 % (0 - 76 mmHg) CO2: AwRR: 2 to 100 BrPM AwRR 20 - 40 s : Refreshing rate 1s Calibration Yearly calibration requested.
Chapter 3 Product Specifications water trap and adult DRYLINE™ sampling line (2.
Chapter 4 Disassembling/Assembling & Troubleshooting Chapter 4 Disassembling/Assembling & Troubleshooting 4.1 PM-9000 Express Disassembling/Assembling 4.1.1 Exploded View of PM-9000 Express Figure 4-1 Exploded view of PM-9000 Express NO Material code Part & Specification Quantity 1 9201-30-35947 Front cover assembly 1 2 9210-30-30181 Back plate assembly 1 3 M04-000305--- Cross-head self-tapping screw 3*12 5 4 M02-000802--- Flat washer GB97.
Chapter 4 Disassembling/Assembling & Troubleshooting 4.1.2 PM-9000 Express Display (TFT Display) Assembly Figure 4-2 PM-9000 Express display (TFT display) assembly NO Standard Name & Specification 1 M04-004015--- Cross-head screw M3×8 Quantity 4 2 0010-10-12271 LG display LB121S02 1 3 9210-20-30180 12.
Chapter 4 Disassembling/Assembling & Troubleshooting 4.1.
Chapter 4 Disassembling/Assembling & Troubleshooting 20 9201-20-36012 Power PCB insulating plate 1 21 9201-30-35901 Lithium battery power PCB 1 22 9210-30-30163 Socket assembly 1 23 9201-30-35908 Microstream CO2 adapter board 1 24 9201-30-35955 Microstream CO2 module 1 25 9201-20-35928 Mounting plate of microstream CO2 1 module 4.1.
Chapter 4 Disassembling/Assembling & Troubleshooting 11 9200-20-10472 Button 12 9200-20-10473 Button backer 13 9201-20-36031 Connector 14 M04-051004--- Cross-head self-tapping screw 2.6*6 15 9201-30-35912 9201 button panel 4.1.
Chapter 4 Disassembling/Assembling & Troubleshooting 12 9200-20-10511 Foot plate 1 13 M04-000501--- Stainless steel nut GB6170MS 14 9201-30-35978 Fan assembly 4.1.
Chapter 4 Disassembling/Assembling & Troubleshooting 4.2.
Chapter 4 Disassembling/Assembling & Troubleshooting 4.2.2 White Screen & Other Abnormal Screen In case of faults causing white screen or other abnormal screens, ■ Check whether the LCD connection wires are in good contact; ■ Replace the LCD connection wires, or replace the LCD if necessary; ■ Replace the main control board if the fault still exists. 4.2.3 Encoder Faults 1.
Chapter 5 Test and Material List Chapter 5 Test and Material List 5.1 Test Procedure 5.1.1 Connection and Checking Connect the simulators, power supply and test fixture properly to the PM-9000 Express patient monitor, and power it on. Then, the patient monitor displays the start-up screen on the TFT screen and enters the system screen. 5.1.2 Functions of Buttons Press every button on the button panel to check their functions as specified in PM-9000 Express Operation Manual.
Chapter 5 Test and Material List the TEMP parameter should be 35±0.1℃, 37±0.1℃ and 39±0.1℃. 2. CY-F1 probe Select CY-F1 probe from the manufacturer menu, select CY-F1 temperature probe as the test fixture, set the analog resistance to 6.534K, 6.018K and 5.548K. Then the TEMP parameter should be 35±0.1℃, 37±0.1℃ and 39±0.1℃. 5.1.5 NIBP Connect the NIBP simulator, adult cuff and accessories, and then connect the module CUFF and clockwise screw it tightly. 1.
Chapter 5 Test and Material List Set the simulator output to ART wave. Then the screen of the PM-9000 Express should display relevant waveform properly. Unplug the IBP probe. Then the screen should prompt “IBP: Transducer 1 OFF!” and “IBP: Transducer 2 OFF!” Plug the OHMEDA cable to the IBP1 channel. Then the prompting message “IBP: Transducer 1 OFF!” disappears. ② IBP2 test: Plug the IBP cable to the IBP2 channel, and repeat the procedure in Section ①. 5.1.8 CO 1.
Chapter 5 Test and Material List prompts “CO2 too high” on the main screen. When the measured value is lower than the low limit, the patient monitor prompts “CO2 too low”. 5.1.10 Recorder 1. Print the ECG waveform. The recorder should print it normally and clearly. Set the recorder to the fault of lack of paper and abnormal clip. There should be relevant prompting messages on the main screen. When the fault is cleared, the patient monitor should become normal. 2.
Chapter 5 Test and Material List 5.2 NIBP Calibration Figure 5-1 NIBP Calibration Calibration method: Based on the precision of 50mmHg (6.7kPa), increase the pressure step by step. The maximum error at any pressure point within the NIBP measurement range of the patient monitor should be no more than ±3mmHg (±0.4kPa). Decrease the pressure step by step. The maximum error at any pressure point within the NIBP measurement range of the patient monitor should be no more than ±3mmHg (±0.4kPa). 5.
Chapter 5 Test and Material List Figure 5-2 IBP PRESSURE ZERO Zero Calibration of Transducer Select CH1, the system will zero IBP1. Select CH2, the system will zero IBP2. Cautions:( Use the PM-6000 IBP module as a example) Turn off patient stopcock before you start the zero procedure. The transducer must be vented to atmospheric pressure before the zero procedure. The transducer should be placed at the same height level with the heart, approximately mid-axially line.
Chapter 5 Test and Material List Figure 5-4 IBP Calibration Menu Calibrate the transducer: Turn the knob to select the item CH1 CAL VALUE, press and turn the knob to select the pressure value to be calibrated for channel 1. Then turn the knob to select the item CALIBRATE to start calibrating channel 1. Turn the knob to select the item CH2 CAL VALUE, press and turn the knob to select the pressure value to be calibrated for channel 2.
Chapter 5 Test and Material List The Calibration Procedure: 1. Close the stopcock that was open to atmospheric pressure for the zero calibration. 2. Attach the tubing to the sphygmomanometer. 3. Ensure that connection that would lead to patient is off. 4. Connect the 3-way connector to the 3-way stopcock that is not connected to the patient catheter. 5. Open the port of the 3-way stopcock to the sphygmomanometer. . 6.
Chapter 5 Test and Material List Note Neither the mainstream nor the sidestream module can be calibrated. Only the overall performance and accuracy is checked. If the Co2 module fails the tests it should be replaced. Figure 5-7 Factory Maintain Menu Figure 5-8 CO2 check menu 5.5 AG CALIBRATE 5.5.1 AG Check 1、Using T-piece to connect the watertrap and Agent steel bottle well. One of the T-piece ports must be vented to atmospheric pressure.
Chapter 5 Test and Material List 5、Choose other pump rate ‘middle’ or ‘high’,and repeat the previous procedures. (pump rate definition: three pump rate under adult mode: 100/150/200ml/min; neonate: 70/90/110 ml/min) 6、If the accurate over range, please press ‘START CAL’. 5.5.2 AG CALIBRATE (Agent>1.5%, CO2>1.5%, N2O>40%, O2>40% ) 1、Press ‘START CAL’, then input password ‘MINDRAY’ entering ‘CALIBRATE’ menu. Note: Make sure the AG in ‘Measure’ mode not ‘Standby’ mode before you do calibrate.
Chapter 5 Test and Material List 5.
Chapter 6 Maintenance and Cleaning Chapter 6 Maintenance and Cleaning 6.1 Maintenance 6.1.1Checking Before Using ■ Check the patient monitor for mechanical damages; ■ Check all exposed conductors, connectors and accessories; ■ Check all functions that are possibly enabled for the monitored patient, and ensure the device is in good working status. In case of any damage, stop using this patient monitor, and contact biomedical engineers of the hospital or Mindray maintenance engineers. 6.1.
Chapter 6 Maintenance and Cleaning 5. Ethanol 6. Isopropyl alcohol 6.4 Sterilization To avoid the long-time damage to the patient monitor, we recommend you To conduct only sterilization which is considered necessary in your maintenance plan; To clean the patient monitor before the sterilization; To sterilize the patient monitor with specified sterilization agent: Ethylate, and Acetaldehyde.
P/N: 9201-20-35991