5.2 Setting the Auxiliary Functions
5-21
RF-20A/20Axs
5
Setting the Output Mode for Analog Output Connector 2 [ANALOG2 MODE]
RF-20A
Set whether a Chromatopac or a recorder is connected
to analog output connector 2.
^ "4.1.3 Setting the Analog Output Connectors"
P. 4 - 5
"4.2.3 Setting the Analog Output Connectors"
P.4-25
RF-20Axs
Enter the set value from the numeric keys and press
.
: RF-20Axs only
ANALOG2 MODE 0
Input 0 - 1
ANALOG2 MODE 0
Input 0 - 1,8
Measurement
Mode
Set
Value
Output Mode
Single
wavelength
mode
0
A Chromatopac or variable range
recorder is connected to analog
output connector 2.
1
A fixed range recorder is connected
to analog output connector 2.
8
The temperature of the flow cell is
output.
Output range: 0 to 100 °C (0 to
10 mV)
Dual
wavelength
mode
0
Set when a Chromatopac or
variable range recorder is
connected to analog output
connector 2 and the channel 1 data
is recorded by the Chromatopac.
1
Set when a fixed range recorder is
connected to analog output
connector 2 and the channel 1 data
is recorded by the recorder.
2
Set when a Chromatopac or
variable range recorder is
connected to analog output
connector 2 and the channel 2 data
is recorded by the Chromatopac.
3
Set when a fixed range recorder is
connected to analog output
connector 2 and the channel 2 data
is recorded by the recorder.
8
The temperature of the flow cell is
output.
Output range: 0 to 100 °C (0 to
10 mV)