5.3 Setting the VP Functions
The VP functions support the validation of the instrument through the display of instrument information and
check functions.
There are the following four types of VP function.
NOTE
When this instrument is used in the RF-10AXL compatibility mode, some functions may no longer be usable,
and the setting ranges may be restricted.
^ "5.7.3 Notes on Operation" P.5-77
5.3.1 List of VP Functions
The VP functions are listed in the tables below.
^ "5.1.3 VP Function Screen" P.5-8
Product Information Group
Maintenance Information Group
Group Main Function
PRODUCT To display the instrument's serial number, unit name, ROM version and so on.
MAINTENANCE
To display the instrument's operating time, the Xenon lamp's cumulative operating time, and
so on.
VALIDATION
To set the time and date, to check the memory and wavelength accuracy, to run performance
checks by the raman spectrum of water, and so on.
CALIBRATION To set the password, to set the Xenon lamp replacement alert time, and so on.
Name Operation Function Page
SERIAL NUMBER Display To show the serial number of the instrument. P. 5- 3 7
S/W ID: V Display To show the unit name of the instrument, and the ROM version. P. 5 -3 7
Name Operation Function Page
TOTAL OP TIME Display To show the cumulative operating time of the instrument. P. 5 - 3 7
Xe LAMP USED TM Numeric keys
To show the Xenon lamp's replacement alert time, and its
cumulative operating time.
P. 5 -3 8
Xe LAMP USED CT Display To show the ignition count of the Xenon lamp. P. 5 - 38
PART REPLACEMENT Numeric keys To enter the part numbers of parts replaced in maintenance. P. 5 - 3 8
MAINTENANCE LOG Display To show the maintenance log. P. 5 -3 9
OPERATION LOG Display To show the operation log. P. 5 -3 9
ERROR LOG Display To show the error log. P. 5 -4 0