Instruction manual

5.3 Setting the VP Functions
5-43
RF-20A/20Axs
5
5.3.6 Calibration Support Group
This is the group for calibrating this instrument.
NOTE
The instrument is adjusted before leaving the factory. Do
not change values unnecessarily.
Entering/Setting the Password [INPUT PASSWORD]
In order to prevent changes being made unnecessarily, a
password has to be entered to change any of the items
in the calibration group. If the password doesn't match,
you cannot proceed to any setting screen. The password
should be set and changed by the system administrator.
Enter a 5-digit number with the numeric keys and press
.
Be sure to set a 5-digit number. The password on
shipment is [00000].
When the password matches:When the password matches, [WAVE CALIB] is
displayed.
When the password doesn't match:When the password doesn't match, the screen to the
right is displayed and you cannot proceed to any setting
screen.
CALIBRATION
Press func or VP
INPUT PASSWORD
*****
WAVE CALIB
Enter to Calib
INPUT PASSWORD
PASSWORD WRONG