RF-20A/20Axs
Executing Wavelength Calibration [WAVE CALIB]
Performs wavelength calibration and a wavelength
accuracy check automatically.
NOTE
When performing wavelength calibration on an RF-20A,
fit the low-pressure Hg (mercury) lamp before starting
wavelength calibration.
^ "8.7 Performing Wavelength Calibration" P.8-39
"9.1.10 Fitting the Low-Pressure Hg (Mercury)
Lamp (RF-20A Only)" P.9-37
1
Press .
[CELL No.] will be displayed.
2
Enter the cell number set for the instrument with
the numeric keys and press .
The cell number of the flow cell unit provided with
the instrument on shipping is set as [1].
Wavelength calibration will start.
^ "Setting the Cell Number [CELL No.]"
P.5-31
"5.9 Using the Spare Flow Cell Unit /
Optional Cell" P.5-84
"8.3 Disassembling the Flow Cell Unit and
Cleaning/Replacing Each Part" P.8-11
The time required for wavelength calibration is
approximately 30 minutes. To discontinue the
wavelength calibration, press .
^ "8.7 Performing Wavelength Calibration" P.8-39
Setting the Replacement Alert Time for the Xenon Lamp [Xe TIME]
Set the replacement alert time for the Xenon lamp.
Enter the set value from the numeric keys and press
.
The default value is [2000] hours.
WAVE CALIB
Enter to Calib
CELL No. 1
Input 1 - 10
Xe TIME 2000
Input Alert Lv
Setting Range (Units: hours)
1 to 2000