5. Application Operation
5-54
RF-20A/20Axs
5.4 Creating Time Programs
This instrument sets parameters such as wavelength, response, output range and so on, and these are
executed by a time program. The created time program is saved when the power is turned OFF.
5.4.1 List of Commands That Can Be Used in Time Programs
Command Operation Function
Setting Range
Page
RF-20A/20Axs
In RF-10AXL
Compatibility
Mode
XCH1 Numeric keys
To set the excitation wavelength for
channel 1.
0, 200 to 900 nm P. 5 - 1 6
XCH2 Numeric keys
To set the excitation wavelength for
channel 2 (only in the dual
wavelength mode).
200 to 900 nm Cannot be set P. 5 -1 6
MCH1 Numeric keys
To set the emission wavelength for
channel 1.
0, 200 to 900 nm P. 5 - 1 6
MCH2 Numeric keys
To set the emission wavelength for
channel 2 (only in the dual
wavelength mode).
200 to 900 nm Cannot be set P. 5 -1 6
ZERO key To zero-adjust the recorder. P. 5 -2 4
MARK key To set marking at recorder output. P. 5 -2 5
RNG1 Numeric keys
To set the output range for analog
output connector 1.
0 to 9 Cannot be set P. 5 -2 2
RNG2 Numeric keys
To set the output range for analog
output connector 2.
0 to 9 P. 5 - 23
RESP Numeric keys Set the response.
0: (without filter)
1: 0.05 sec
2: 0.1 sec
3: 0.5 sec
4: 1.0 sec
5: 1.5 sec
6: 3.0 sec
7: 6.0 sec
8: 8.0 sec
9: 10.0 sec
10: 2.0 sec
1: 0.1 sec
2: 0.5 sec
3: 1.5 sec
4: 3.0 sec
P. 5 -1 8
SENS Numeric keys To set the sensitivity.
1: HIGH
2: MED
3: LOW
P. 5 -1 9
GAIN Numeric keys To set the gain.
1: × 1
2: × 4
3: × 16
P. 5 -1 9