Instruction manual
5.4 Creating Time Programs
5-57
RF-20A/20Axs
5
Example of Creation of a Time Program
Shown here is an example where spectrum scanning is executed 5 minutes after the time program starts, the
data is saved in file 2, and the time program is terminated after 10 minutes.
NOTE
• When multiple steps are set they are automatically re-ordered, so there is no need to set them in
chronological sequence.
• Except when running the time program endlessly, set [STOP] at the end of the program.
• Pressing at command selection displays the previous command.
1
Press .
The initial screen will be displayed.
2
Press .
The screen showing the number of time program
steps will be displayed.
3
Press .
The screen for setting [TIME] will be displayed.
4
Press and .
Step TIME (Minutes) Command Set Value
1 5.0 SCAN
2 (specifies the file number in
which the data is saved)
210 STOP −
EX200nm EM300nm
Xe 1000.00
TIME PROGRAM
0 Used 32 Left
TIME FUNC VALUE
Input 0.01 - 999