Instruction manual

5.4 Creating Time Programs
5-57
RF-20A/20Axs
5
Example of Creation of a Time Program
Shown here is an example where spectrum scanning is executed 5 minutes after the time program starts, the
data is saved in file 2, and the time program is terminated after 10 minutes.
NOTE
When multiple steps are set they are automatically re-ordered, so there is no need to set them in
chronological sequence.
Except when running the time program endlessly, set [STOP] at the end of the program.
Pressing at command selection displays the previous command.
1
Press .
The initial screen will be displayed.
2
Press .
The screen showing the number of time program
steps will be displayed.
3
Press .
The screen for setting [TIME] will be displayed.
4
Press and .
Step TIME (Minutes) Command Set Value
1 5.0 SCAN
2 (specifies the file number in
which the data is saved)
210 STOP
EX200nm EM300nm
Xe 1000.00
TIME PROGRAM
0 Used 32 Left
TIME FUNC VALUE
Input 0.01 - 999