Instruction manual

5.4 Creating Time Programs
5-59
RF-20A/20Axs
5
5.4.3 Setting the Loop Count of the Program [LOOP]
By setting [LOOP], the time program can be repeated
the set number of times.
When the settings in the table to the right are made,
steps 1 and 2 are repeated 3 times in a 30-minute cycle.
Set [LOOP] at the very end of the program.
NOTE
Values of up to 255 can be set for the [LOOP]
command. Note that when [0] is set, the program will
be repeated 256 times.
If any time program is set after the [LOOP] command,
it will not be executed. The time program stops at the
point where the [LOOP] command ends.
5.4.4 Deleting Steps
Delete unnecessary steps from a time program.
1
Display the step that you want to delete.
If deleting the second or a later step, repeatedly
press until the step that you want to
delete appears.
2
Press .
The displayed step is deleted.
If the next step is set, that step is displayed.
30.00 LOOP 3
Input 0 - 255
Step TIME FUNC VALUE
1 15.00 XCH1 210
2 20.00 XCH1 220
3 30.00 LOOP 3
5.00 SCAN 2
Input 0 - 2