Instruction manual

5. Application Operation
5-60
RF-20A/20Axs
5.4.5 Starting a Time Program
There are the following two methods for starting a time
program.
Press .
Input contact signals to the external input/output
terminals.
^ "5.8 Connection to External Input/Output
Terminals" P.5-81
The time program indicator on the display panel will
come on and the time program will start.
NOTE
If set values are changed during the execution of a time
program, the values after the changes will be applied
only up until the time program ends. After the time
program has ended, the settings made before the
execution of the time program will be reinstated.
5.4.6 Stopping a Time Program [STOP]
There are the following three methods for stopping a time program.
Press during the execution of the time program to forcibly end it.
Send contact signals to the external input/output terminals during the execution of the time program to
forcibly end it.
^ "5.8 Connection to External Input/Output Terminals" P.5-81
Insert a [STOP] command in the time program.
^ "Example of Creation of a Time Program" P.5-57
Time program indicator
run
run