5.5 Measuring in the Spectrum Scanning Mode
5-65
RF-20A/20Axs
5
5.5.5 Setting the Excitation Wavelength and Emission Wavelength
Set the excitation and emission wavelengths for spectrum scanning. If [1] (excitation scanning) was set in
"5.5.3 Setting the Scan Type" P.5-63 the emission wavelength is set, and if [2] (emission scanning) was set
the excitation wavelength is set.
The setting ranges for the wavelengths are as follows.
NOTE
When this setting is changed, the wavelength for channel 1 in the parameter settings group is also changed.
^ "Setting the Wavelength for Channel 1 [ch1]" P.5-16
1
Press repeatedly until [ch1] in the control
settings group is displayed.
In the initial status, the screen shows the value
set in "Setting the Wavelength for Channel 1
[ch1]" P.5-16.
2
Enter the wavelength with the numeric keys and
press .
The wavelengths will be set.
If [1] was set for the scan type in [SPC TYPE]
setting the emission wavelength is set, and if [2]
was set the excitation wavelength is set.
^ "5.5.3 Setting the Scan Type" P.5-63
Wavelength Setting Range (Units: nm)
EX (excitation wavelength) 0, 200 to 900
EM (emission wavelength) 0, 200 to 900
ch1 EX3 50 EM450
Input 0,2 0 0-900