5.5.8 Setting Gain
This is the procedure for setting gain. By combining the setting of sensitivity and gain, you can set the
measuring range. The settings for gain are shown below.
The combinations of sensitivity and gain are shown below.
1
Press repeatedly until [GAIN] in the
parameter settings group is displayed.
2
Enter the set value from the numeric keys and
press .
This sets the gain.
Set Value Gain
1 × 1
2 × 4
3 × 16
Sensitivity Gain Sensitivity Magnifications
3 (LOW)
1 Approx. × 1
2 Approx. × 4
3 Approx. × 16
2 (MED)
1 Approx. × 32
2 Approx. × 128
3 Approx. × 512
1 (HI)
1 Approx. × 1024
2 Approx. × 4096
3 Approx. × 16384
GAIN 2