Instruction manual

Precautions for Mobile Phase Selection and Use
XV
RF-20A/20Axs
Precautions for Mobile Phase Selection and
Use
!CAUTION
If PEEK resin parts are used in the plumbing, do not use the following mobile phases.
These mobile phases weaken the PEEK resin, which could result in cracked plumbing
and mobile phase leaks:
Concentrated sulfuric acid, concentrated nitric acid, dichloroacetic acid, acetone, tetrahydrofuran
(THF), dichloromethane, chloroform, dimethyl sulfoxide (DMSO), fluorine organic solvents such as
hexafluoroisopropanol (HFIP)
Note: There is no problem with temporarily using a low-concentration aqueous solution with an acetone
concentration of 0.5 % or less, e.g. for the purpose of checking the performance of the gradient.
NOTE
Use only HPLC grade or comparable mobile phase, and filter it with a filter of 0.45 μm mesh or finer
before use to remove particulates and foreign matter.
Halogen ions can corrode the stainless steel material (SUS316L) used in the plumbing, so avoid, as
much as possible, mobile phases that contain halogen ions - such as KCI, NaCI and NH
4
Cl - or mobile
phases that generate halogen ions in certain reactions. If such mobile phases must be used, clean all
flow lines thoroughly with water immediately after analysis.
When SPD or a similar UV detector is used for high-sensitivity analysis, be sure to use HPLC grade
mobile phases that have a low absorptivity of UV rays.
Always degas the mobile phase, as air bubbles may tend to form during solvent mixing or during
temperature or pressure changes. Air bubbles may cause pump malfunctions and detector signal
noise.
Understand the properties, including viscosities, of the mobile phase.
^ "9.5 Mobile Phase Characteristics" P.9-48