Instruction manual

5.5 Measuring in the Spectrum Scanning Mode
5-73
RF-20A/20Axs
5
5.5.12 Outputting Spectrum Data
This is the procedure for outputting the read spectrum data. Before output, set the speed at which the
spectrum data is to be output.
Details of the output speed settings are given below.
1
Press repeatedly until [PLOT SPD] in the
control settings group is displayed.
2
Enter the plot speed setting value from the
numeric keys and press .
3
Press or until [SCAN FILE] is
displayed.
4
Enter the file number of the file to be output to with
the numeric keys and press .
5
Press or until [SPC PLOT] is
displayed.
6
Press .
Output will start.
Set Value Plot Speed (Units: nm/sec)
11
25
310
PLOT SPD 1
1:1 2:5 3:10nm/s
SCAN FILE 0
0:BG 1:S1 2:S2
SPC PLOT
Enter to Plot