Instruction manual

5. Application Operation
5-74
RF-20A/20Axs
To Stop Output Part Way Through
Press while the screen to the right is displayed.
Fig. 5.2
Example of output
The output time, T, is determined as follows.
When wavelengths from 200 nm to 350 nm are output at a plot speed of 5 nm/sec ([PLOT SPD] = [2]):
NOTE
Only a file generated by spectrum scanning can output data. If there is no data in the file, the message
[DATA NOT EXIST] will be displayed. Note that output will also not be possible if there is no data in the
background file.
^ "5.5.10 Setting the Scan File" P.5-70
Spectrum data is erased when the power to the instrument is turned OFF.
If the scan conditions (start wavelength, end wavelength, scan speed, etc.) are changed from those under
which the scan was executed, correct output may not be possible.
SPC PLOT
PLOTTIING
Marks are inserted at the start and the end.
Zero position
T =
(SCAN END - SCAN BGN)
Plot speed (1, 5, 10 nm/sec)
(350 - 200)
5 nm/sec
T = = 30 sec