5.7 Connecting to an SCL-10Avp System Controller
5-77
RF-20A/20Axs
5
5.7.3 Notes on Operation
Restrictions in the RF-10AXL Compatibility Mode
When this instrument is used while connected to an SCL-10Avp, it operates as an RF-10AXL and the
functions are therefore restricted in the following way.
The dual wavelength mode cannot be used.
The EVENT2 terminal in the external input/output terminals cannot be used.
The temperature of the flow cell cannot be set from the SCL-10Avp. Set the temperature at the instrument
itself.
^ "Setting the Flow Cell Temperature [CELL TEMP] (RF-20Axs Only)" P.5-17
The settings of analog output connectors 1 and 2 ([ANALOG1 MODE], [ANALOG2 MODE]) are fixed as
follows.
^ "Setting the Output Mode for Analog Output Connector 1 [ANALOG1 MODE]" P.5-20
"Setting the Output Mode for Analog Output Connector 2 [ANALOG2 MODE]" P.5-21
The setting range for the response is as follows.
^ "Setting the Response [RESPONSE]" P.5-18
The setting ranges of the commands that can be used in time programs are restricted.
^ "5.4.1 List of Commands That Can Be Used in Time Programs" P.5-54
Connector Set Value Output Mode
Analog output connector 1 0 INTEGRATOR (Connect a Chromatopac or a variable range recorder.)
Analog output connector 2 1 RECORDER (Connect a fixed range recorder.)
Set Value Time Constant (Units: sec)
10.1
20.5
31.5
43.0