Instruction manual

6.2 Dealing with Error Messages
6-13
RF-20A/20Axs
6
ERR FILE TYPE
Warning
Cause: Displayed when the spectrum types or
wavelength ranges of the data in file number [0]
and the data to be output are different when
[SPC PLOT] is executed and the scan data is
output.
(File type error)
Action: Rescan so that the spectrum type or
wavelength range of the scan data to be output
matches that of the data in file number [0], then
execute [SPC PLOT] again.
^ "Outputting Spectrum Data [SPC PLOT]"
P.5-28
LAMP NOT LIT
Warning
Cause: Displayed if or is pressed while
the Xenon lamp is off.
(Warning when lamp not lit)
Action: Set [1] (ON) for [LAMP] in the parameter
settings group.
^ "Setting the Lamp ON/OFF Status
[LAMP]" P.5-17
SENSOR NO GOOD
Warning
Cause: Displayed when a leak could not be detected on
execution of [LEAK SENSOR TEST].
(Warning on leak sensor test)
Action: Calibrate the leak sensor and change its
actuation level (threshold value).
^ "Calibrating the Leak Sensor [L-CAL]"
P.5-45
"Setting the Leak Sensor Actuation Level
[LEAK THR]" P.5-45
SELECT Xe SINGLE
Warning
Cause: Displayed when [EX CHECK] or [EM CHECK]
is executed when the measurement mode is not
the single wavelength mode while the Xenon
lamp is on.
(Wavelength mode warning)
Action: Set [1] (single wavelength mode) for [λ MODE]
in the parameter settings group.
^ "Setting the Measurement Mode [λ
MODE]" P.5-17
Error Message Type Cause and Action