6. Troubleshooting
6-14
RF-20A/20Axs
LOW SET TEMP
Warning
Cause: Displayed if the room temperature exceeds the
set temperature for the flow cell by 10 °C or
more.
(Flow cell temperature control error)
RF-20Axs only
Action: Set a flow cell temperature that is within 10 °C
of the room temperature. If the warning is still
displayed, turn the power OFF and contact your
Shimadzu representative.
^ "Setting the Flow Cell Temperature
[CELL TEMP] (RF-20Axs Only)" P.5-17
CLOSED KEY
Warning
Cause: Displayed on pressing an operation key when
the setting to prohibit key input has been made
with [KEY CLOSE].
(Keys locked)
Action: Press while holding down to
cancel the prohibition on key input.
^ "Prohibiting Key Input [KEY CLOSE]"
P.5-29
NO CAL DATA
Warning
Cause: Displayed on input of a cell number for which
wavelength calibration has not been performed
([CELL No.] in the system settings group)
("Wavelength not calibrated" alarm)
Action: Input a cell number for which wavelength
calibration has been performed. Or, carry out
wavelength calibration for the cell number that
you attempted to input.
^ "Setting the Cell Number [CELL No.]"
P.5-31
"8.2 Inspection and Simple Washing of
the Cell" P.8-5
NOT LOCAL MODE
Warning
Cause: Displayed on attempting to execute a function
that can only be used when the local mode is
[Local] while the local mode is set to [Remote].
("Not local mode" alarm)
Action: Set [1] (Local) for [LOCAL] in the system
settings group.
^ "Setting the Local Mode [LOCAL]" P.5-29
Error Message Type Cause and Action