7.5 Detector Validation
7-7
RF-20A/20Axs
7
7.5 Detector Validation
7.5.1 Check Points
The check points for detector validation are listed below.
Section Check Points Overview
7.5.2
Initialization check and ROM,
RAM self diagnosis
For the RF-20Axs:
Check the operation of the display, LEDs and drive section, check for
correction operation of the wavelength accuracy check using the
emission line of the low-pressure Hg (mercury) lamp, and check that the
memory (ROM/RAM) is normal.
For the RF-20A:
Check the operation of the display, LEDs and drive section, and check
that the memory (ROM/RAM) is normal.
7.5.3 Firmware version check Check the version of the firmware.
7.5.4
Check on cumulative operating
time of Xenon lamp
Check the cumulative operating time of the Xenon lamp.
7.5.5 Wavelength accuracy check
Check the wavelength accuracy using the emission line of the low-
pressure Hg (mercury) lamp.
7.5.6 Performance inspection Inspect a performance using the raman spectrum of water.
7.5.7 Leak sensor test Check the operation of the leak sensor.