Instruction manual

7.5 Detector Validation
7-17
RF-20A/20Axs
7
[EX CHECK]
10
Repeatedly press until [EX CHECK] or
[EM CHECK] appears.
[EM CHECK]
NOTE
For a wavelength accuracy check at the excitation side,
execute [EX CHECK], and for a wavelength accuracy
check at the emission side, execute [EM CHECK].
11
Press .
[EX CHECK] or [EM CHECK] starts.
During wavelength accuracy check the screen to
the right is displayed.
The wavelength that is being checked is indicated
at [***]. When the wavelength accuracy check
ends, the wavelength accuracy is displayed on the
screen.
EX CHECK
Enter to Start
EM CHECK
Enter to Start
EX CHECK
CHECKING EX ***
EM CHECK
CHECKING EM ***