7. Hardware Validation
7-20
RF-20A/20Axs
NOTE
After the wavelength check has ended, be sure to
remove the low-pressure Hg (mercury) lamp, and fit the
Xenon lamp as it was fitted before.
After the wavelength accuracy inspection, perform the operation below.
When neither a system controller nor LCsolution is used:
1
Turn the power to the instrument OFF.
2
Remove the low-pressure Hg (mercury) lamp, and fit the Xenon lamp as it was originally fitted.
When the system controller is connected:
1
Turn the power to the instrument and the system controller OFF.
2
Remove the low-pressure Hg (mercury) lamp and fit the Xenon lamp as it was originally fitted.
3
Turn the power to the instrument ON and set [0] for [LOCAL] in the system settings group.
4
Restart the system controller.
When LCsolution is used:
1
Exit LCsolution.
2
Turn the power to the instrument and the system controller OFF.
3
Remove the low-pressure Hg (mercury) lamp and fit the Xenon lamp as it was originally fitted.
4
Turn the power to the instrument ON and set [0] for [LOCAL] in the system settings group.
5
After restarting the system controller, start LCsolution.
CHECK CRITERIA: For both excitation side and emission side
Wavelength accuracy at 254 nm: Within ± 2.0 nm
Wavelength accuracy at 507 nm: Within ± 2.0 nm