7. Hardware Validation
7-24
RF-20A/20Axs
6
Press .
The instrument is automatically given the following
settings.
After setting has finished, automatic zeroing is
executed and the screen to the right is displayed.
7
Set a Chromatopac as follows.
8
Check the input signal level of the Chromatopac.
NOTE
Check whether the input signal is within the range -1000
to +5000 uV. If it is outside this range, press to
adjust the zero position, then check the input signal level
again.
9
Execute auto zeroing of the Chromatopac and
start the Chromatopac record.
Item Set Set Value
EX (excitation wavelength) 350 nm
EM (emission wavelength) 450 nm
GAIN 2 (× 4)
SPC TYPE 2 (emission scanning)
Start wavelength for EM SCAN 350 nm
End wavelength for EM SCAN 450 nm
SENS 2 (MED)
RESPONSE 5 (1.5 sec)
ANALOG1 MODE 0 (Chromatopac)
S/N CHECK ENTER
Enter to Start
Item Set Set Value
ATTEN 7 (128 mV/full scale)
SPEED 50 mm/min