7.5 Detector Validation
7-29
RF-20A/20Axs
7
6
Repeatedly press until [S/N CHECK]
appears.
7
Press .
The instrument is automatically given the following
settings.
After setting has finished, automatic zeroing is
executed and the screen to the right is displayed.
8
Press .
Spectrum scanning for the emission wavelength
takes place automatically.
During spectrum scanning the screen to the right
is displayed.
S/N CHECK
Enter to Start
Item Set Set Value
EX (excitation wavelength) 350 nm
EM (emission wavelength) 450 nm
GAIN 2 (× 4)
SPC TYPE 2 (emission scanning)
Start wavelength for EM SCAN 350 nm
End wavelength for EM SCAN 450 nm
SENS 2 (MED)
RESPONSE 5 (1.5 sec)
ANALOG1 MODE 0 (Chromatopac)
S/N CHECK ENTER
Enter to Start
S/N CHECK1000.00
EX:350 EM:350
Intensity of the emission