Instruction manual

7. Hardware Validation
7-34
RF-20A/20Axs
4
After about 10 seconds, press .
When a leak is detected:When a leak is detected, [CHECK GOOD] is
displayed.
When no leak is detected:When no leak is detected, [SENSOR NO GOOD]
is displayed.
5
Press .
The result display is cleared.
If the result is [SENSOR NO GOOD],
automatically calibrate the leak sensor and
change its actuation level by following the
sections on [L-CAL] and [LEAK THR].
^ "Calibrating the Leak Sensor [L-CAL]"
P.5-45
"Setting the Leak Sensor Actuation Level
[LEAK THR]" P.5-45
LEAK SENSOR TEST
CHECK GOOD
LEAK SENSOR TEST
SENSOR NO GOOD
CHECK CRITERIA: [CHECK GOOD] is displayed on the screen.