Instruction manual

7. Hardware Validation
7-46
RF-20A/20Axs
7.7 If Validation Fails
Should the system fail to satisfy any of the system validation check criteria, or should a component fail to
satisfy any of the component validation check criteria, proceed as follows.
Check whether any consumable parts have reached the end of their service life:
The cause of failure to satisfy check criteria could be a consumable part that is no longer usable. Check
consumable parts and replace them if necessary.
Perform troubleshooting:
It is possible that some minor problem (such as air bubbles) has caused the system to fail the criteria.
Perform troubleshooting to check for such problems, and take action to eliminate any problems found.
For troubleshooting procedures for individual system components, refer to the applicable instruction
manuals.
If a cause cannot be determined, contact your Shimadzu representative:
If you are unable to determine the cause of the failure, or if you are unclear about troubleshooting or
corrective action procedures, contact your Shimadzu representative.