Instruction manual

8. Maintenance
8-6
RF-20A/20Axs
Inspecting the Cell
1
Turn the power to the instrument OFF and unplug
the power plug.
2
Open the front cover and the key panel.
NOTE
In order to prevent the front cover falling off, remove it
from the instrument before starting the work.
3
Unscrew the coupling fixing screw and remove the
inlet and outlet couplings 1.6-0.8C.
4
Unscrew the cell housing screw and remove the
flow cell unit.
Fig. 8.3
5
Pump with the flow line tubing still connected.
6
With the mobile phase flowing, look inside the flow
cell unit through the lens, and check for air
bubbles and dirt.
If there are any air bubbles or is any dirt, clean
the cell.
^ "8.2.2 Simple Cleaning of the Cell" P.8-8
Fig. 8.4
Cell housing
screw
Coupling
fixing screw
Flow cell
unit
Couplings
1.6-0.8C
Lens ASSY
To the waste
container
From the column