8.2 Inspection and Simple Washing of the Cell
8-7
RF-20A/20Axs
8
Fitting the Flow Cell Unit
Fit the flow cell unit to the instrument by following the
procedure below.
NOTE
Ensure that no air bubbles get into the flow line during
fitting.
1
Fit the flow cell unit with the lens ASSY facing
upward.
2
Screw in the cell housing screw.
Fig. 8.5
3
Secure the inlet and outlet couplings 1.6-0.8C with
the coupling fixing screw.
Fig. 8.6
4
Close the front cover and key panel.
Lens ASSY
Cell housing
screw
Flow cell unit Excitation inlet
Couplings
1.6-0.8C
Coupling
fixing screw
Inlet coupling 1.6-0.8C
Outlet coupling