Instruction manual

8.3 Disassembling the Flow Cell Unit and Cleaning/Replacing Each Part
8-23
RF-20A/20Axs
8
Fitting the Emission Mirror
NOTE
Fit the emission mirror so that its convex side faces
toward the bottom face of the cell housing.
Tighten the emission mirror fixing screw lightly so as
not to damage the emission mirror.
RF-20A
1
Fit the emission mirror to the cell housing by
following the procedure for removal in reverse.
Fig. 8.32
RF-20A
XS
Fig. 8.33
Cell housing
Bottom face of the
cell housing
Packing
Emission
mirror
Spacer
Fit the emission mirror with its convex side down.
Tighten the emission mirror fixing screw while the
spacer is pressed against the emission mirror.
Fixing
screw for
emission
mirror
Cell housing
Bottom face of the
cell housing
Packing
Emission
mirror
Fixing screw
for emission
mirror
Fit the emission mirror with its
convex side down.