Instruction manual

8.3 Disassembling the Flow Cell Unit and Cleaning/Replacing Each Part
8-25
RF-20A/20Axs
8
1
Fit the excitation lens in the cell housing by
following the procedure for removal in reverse.
Fig. 8.35
Excitation
lens
Fixing screw for
excitation lens
Front face of cell housing
Fit the excitation lens with its convex side facing
outward.