8.3 Disassembling the Flow Cell Unit and Cleaning/Replacing Each Part
8-27
RF-20A/20Axs
8
Fitting the Excitation Mirror
NOTE
Fit the excitation mirror with its concave side facing
toward the inside of the cell housing.
When fitting the fixing screw for the excitation mirror,
do not tighten it any more strongly after it has made
contact with the spacer.
1
Fit the excitation mirror in the cell housing by
following the procedure for removal in reverse.
Fig. 8.37
8.3.7 Assembling and Fitting the Flow Cell Unit
Check After Assembly
After finishing cleaning or replacing each of the parts,
reassemble the flow cell unit.
Before fitting the flow cell unit in the instrument, feed
mobile phase into the flow cell with the syringe and
check that there are no liquid leaks. Lead the outlet side
tube into the waste container.
1
Reassemble the flow cell unit by following the
procedure for disassembly in reverse.
^ "8.3.1 Disassembling the Flow Cell Unit"
P.8-14
Fixing screw for
excitation mirror
Spacer
Excitation
mirror
Front face of cell housing
Fit the excitation mirror with its concave side facing
inward.