8.4 Inspecting/Replacing the Xenon Lamp
8-31
RF-20A/20Axs
8
8.4.1 Replacing the Xenon Lamp
Parts used
1
Turn the power to the instrument OFF, unplug the
power plug, and wait until the Xenon lamp cools.
2
Open the front cover and the key panel.
NOTE
In order to prevent the front cover falling off, remove it
from the instrument before starting the work.
3
Unscrew the knurled screw (white) securing lamp
cover 1, then remove lamp cover 1 and the filter.
The knurled screw (white) is not detachable
from lamp cover 1.
Fig. 8.42
4
Remove the screw securing lamp cover 2, then
remove lamp cover 2.
Fig. 8.43
Part Name Part Type Part No.
Xenon lamp Consumable part 228-51511-95
Knurled
screw
Lamp
cover 1
Filter
Screw Lamp cover 2