8. Maintenance
8-44
RF-20A/20Axs
5
Using the pump, pump water into the flow cell
(flow rate: 1 mL/min).
^ "8.2 Inspection and Simple Washing of the
Cell" P.8-5
6
Check that there are no air bubbles inside the
flow cell, then fit the flow cell to the instrument.
7
Turn the power to the instrument ON.
8
With the initial screen displayed, repeatedly press
until [CALIBRATION] appears.
9
Press .
[INPUT PASSWORD] will be displayed.
10
Enter the password.
When the password matches, [WAVE CALIB] is
displayed.
The default value is [00000].
If the password entered is incorrect, the screen
to the right is displayed.
In this case, check the password and enter it
again.
11
Press .
[CELL No.] will be displayed.
CALIBRATION
Press func or VP
INPUT PASSWORD
*****
WAVE CALIB
Enter to Calib
INPUT PASSWORD
PASSWORD WRONG
CELL No. 1
Input 1 - 10