8.7 Performing Wavelength Calibration
8-47
RF-20A/20Axs
8
When LCsolution is used:
1
Exit LCsolution.
2
Turn the power to the instrument and the system controller OFF.
3
Remove the low-pressure Hg (mercury) lamp and fit the Xenon lamp as it was originally fitted.
4
Turn the power to the instrument ON and set [0] for [LOCAL] in the system settings group.
5
After restarting the system controller, start LCsolution.
NOTE
Before starting analysis when using a system controller or LCsolution, be sure to set [0] for [LOCAL] in the
system settings group, then restart the system controller and LCsolution.