Instruction manual

9.1 Installation
9-3
RF-20A/20Axs
9
Required Installation Space
Typical system configurations and required installation spaces are shown in the figures below.
Fig. 9.1
!CAUTION
The weight of this instrument is 16 kg (RF-20A) or 18 kg (RF-20Axs). During installation,
consider the entire weight combined with other LC components.
The lab table on which this instrument is installed should be strong enough to support the total weight of
the LC system. It should be level, stable and have depth of at least 600 mm.
Otherwise, the instrument could tip over or fall off the table.
Keep at least 100 mm between the rear of the instrument and the wall.
This allows for sufficient air circulation to provide cooling and prevent the instrument from overheating
and impairing the performance.
System 1 (with manual injector)
System 2 (with autosampler)
(Units: mm)