Instruction manual

9. Technical Information
9-4
RF-20A/20Axs
9.1.2 Installation
Removing the Shipping Screws
In order to prevent shock during transportation, the instrument is fixed with the shipping screws. Remove
these screws prior to installation.
NOTE
When the instrument is used without removing shipping screws, noise due to vibration may be heard during
operation.
1
Loosen and remove the shipping screws (with
washers, painted red).
^ "2.3 Right Side and Base Panel" P.2-4
Fig. 9.2
Installation
The instrument is designed for stacking with other Shimadzu HPLC components.
^ "9.4 Introduction to the HPLC System" P.9-45
NOTE
In order to ensure detection with the highest sensitivity and accuracy, install the instrument close to the
column.
It is the general practice to install it above the column oven.
Fig. 9.3
Shipping screws
(painted red)
!CAUTION
Take care to avoid pinching your fingers
between the components.
When the LC-20A series components are stacked
on each other, the clearance between the
components is only 5 mm.