Instruction manual

9. Technical Information
9-6
RF-20A/20Axs
9.1.3 Power Supply Connection
The following table shows the power supply voltage, power consumption, and frequency.
RF-20A
RF-20Axs
Verify that the power outlet to be used for connection has sufficient capacity. If the capacity is insufficient, a
power outage or voltage drop can occur, affecting not only this instrument, but other equipment connected to
the same power supply.
Fig. 9.5
Part No. Power Supply Voltage
Power
Consumption
Frequency
228-45147-41 AC100-120 V (100-120 V~)
400 VA 50/60 Hz228-45147-42 AC100-120 V (100-120 V~)
228-45147-48 AC220-240 V (220-240 V~)
Part No. Power Supply Voltage
Power
Consumption
Frequency
228-45148-41 AC100-120 V (100-120 V~)
400 VA 50/60 Hz228-45148-42 AC100-120 V (100-120 V~)
228-45148-48 AC220-240 V (220-240 V~)
Voltage
indication
!WARNING
The power supply voltage is indicated on the
rear face of the instrument. Be sure to
connect a suitable power supply.
If a power supply voltage other than the one
indicated is used, it will cause fire, electric shock
or instrument failure.
!CAUTION
Fit the Xenon lamp before turning the power switch ON.
The instrument is delivered with the Xenon lamp removed.
When turning the power switch ON, be sure to check that the Xenon lamp is fitted.
When the Xenon lamp is turned on, a high voltage of around 30 kV is applied to the terminal of the
lamp. If the lamp is not fitted correctly at the time of ignition, the instrument may be damaged.
^"9.1.9 Fitting the Xenon Lamp" P.9-33