Instruction manual

9.1 Installation
9-19
RF-20A/20Axs
9
Second unit from bottom:
NOTE
Leaks from the column oven are drained separately (refer to the column oven instruction manual). If any units
are installed on top of the column oven, carry out the same procedure as described in "Installation on top of
the column oven:" P.9-20.
1
Fit the drain adapter at the position shown in the
illustration, and place it on the lower unit.
Fig. 9.27
Cross-section of connection parts
2
The drain adapter connects the drain outlet to the
leakage hole of the lower unit.
3
Pour some water onto a spot near the drain outlet
of the top unit, and verify that the water flows to
the waste container.
Fig. 9.28
Drain adapter
45°
Upper unit
Drain
adapter
Lower unit
Drain outlet
Leakage hole