Instruction manual

9. Technical Information
9-24
RF-20A/20Axs
2
Install a male nut (with long bushing) and ferrule
on one end of each of the two waste liquid tubing
sections. Then attach the tubing and ferrules at
ports 5 and 6 of the manual injector. Tighten the
nuts.
3
Unscrew and remove the vial cap.
4
Route the other ends of the waste liquid tubing
through the tubing opening and into the vial.
Fig. 9.35
NOTE
To prevent liquid from flowing out due to the siphon
effect, position the ends of the waste liquid tubing level
with the needle port.
Fig. 9.36
NOTE
When a manual indicator is fitted to the column oven,
pass the waste liquid tubing through the left door so that
it is perpendicular to the door. If the tubing bends, it may
contact the side face, with the result that the left door will
not close.
Fig. 9.37
Tubing hole
Waste liquid tubing
Vial
Waste liquid tubing
Height of
needle port
Contact with
the side face
Left door
Waste liquid
tubings
Manual
injector
Perpendicular