RF-20A/20Axs
4
Fit the low-pressure Hg (mercury) lamp into the
lamp unit so that the slit in the low-pressure Hg
(mercury) lamp holder faces toward the back of
the instrument.
Fit so that the two holes in the low-pressure Hg
(mercury) lamp holder mate with the positioning
pins on the lamp unit.
Fig. 9.58
5
Using the two hexagon spacers that secured the
Xenon lamp to the instrument, secure the low-
pressure Hg (mercury) lamp holder to the lamp
unit.
Fig. 9.59
6
Place the lamp unit in the instrument as it was
before.
NOTE
The PTFE cover on the top face of the lamp unit need
not be returned to its original position.
When a low-pressure Hg (mercury) lamp has been
fitted, it is no longer possible to fit the filters, lamp
cover 1 or lamp cover 2.