Instruction manual

4. Basic Operation
4-4
RF-20A/20Axs
3
Press .
[ch1] will be displayed.
In the initial status, the excitation wavelength can
be entered.
4
Enter the excitation wavelength with the numeric
keys, then press .
This sets the excitation wavelength.
5
Press .
The status in which the emission wavelength can
be entered will be established.
6
Enter the emission wavelength with the numeric
keys, then press .
This sets the emission wavelength.
7
Press twice.
Setting of the excitation wavelength and emission
wavelength ends and you are returned to the initial
screen.
NOTE
This instrument has two analog output connectors: analog output connector 1 and analog output connector 2.
Set the analog output connector according to the equipment you are connecting to.
^ "4.1.3 Setting the Analog Output Connectors" P.4-5
ch1 EX3 50 EM450
Input 0,2 0 0-900
ch1 EX4 00 EM450
Input 0,2 0 0-900
Connected Equipment Remark
When connecting a Chromatopac or a
variable range recorder
Set the output mode for the analog output connectors to INTEGRATOR.
When connecting a fixed range recorder
Set the output mode for the analog output connectors to RECORDER, and
set the output range at the instrument.
^ "4.1.4 Setting the Output Ranges" P.4-7