4.1 Measuring in the Single Wavelength Mode
4-19
RF-20A/20Axs
4
4.1.9 Setting the Flow Cell Temperature (RF-20Axs Only)
Set the temperature of the flow cell.
The fluorescent intensity of the sample varies depending on its temperature. In order to obtain stable analysis
results unaffected by the ambient temperature, the temperature of the sample in the flow cell is fixed at all
times.
NOTE
The guide for the set temperature of the flow cell is the same temperature as the instrument's ambient
temperature.
Since condensation may occur in the vicinity of the flow cell depending on the ambient humidity if the
temperature of the flow cell goes below the flow cell's ambient temperature, it is made impossible to cool the
flow cell more than 10 °C below the ambient temperature. If the ambient temperature rises to more than 10 °C
higher than the set temperature, [LOW SET TEMP] is displayed on the display screen every 60 minutes and it
may not be possible to obtain normal analysis results.
NOTE
The current ambient temperature can be checked at [ROOM TEMP].
^ "Displaying the Ambient Temperature [ROOM TEMP] (RF-20Axs Only)" P.5-33
The setting range for flow cell temperature is shown below.
1
Press .
The initial screen will be displayed.
2
Press .
[PARAMETER] will be displayed.
Set Value Setting
0 The temperature is not adjusted.
4 to 40 °C
The temperature is adjusted to the set
value.
EX200n m EM30 0nm
X e 1000.00
PARAMETER
Enter to Select