4.2 Measuring in the Dual Wavelength Mode
4-23
RF-20A/20Axs
4
2
Press .
[PARAMETER] will be displayed.
3
Press .
[ch1] will be displayed.
In the initial status, the excitation wavelength can
be entered.
4
Enter the excitation wavelength with the numeric
keys, then press .
This sets the excitation wavelength for channel 1.
5
Press .
The status in which the emission wavelength of
channel 1 can be entered will be established.
6
Enter the emission wavelength with the numeric
keys, then press .
This sets the emission wavelength of channel 1.
7
Press twice.
[ch2] will be displayed.
In the initial status, the excitation wavelength can
be entered.
8
Enter the excitation wavelength with the numeric
keys, then press .
This sets the excitation wavelength for channel 2.
9
Press .
The status in which the emission wavelength of
channel 2 can be entered will be established.
PARAMETER
Enter to Select
ch1 EX3 50 EM450
Input 2 0 0-900
ch1 EX4 00 EM450
Input 2 0 0-900
ch2 EX3 50 EM450
Input 2 0 0-900
ch2 EX4 00 EM450
Input 2 0 0-900