Instruction manual

4. Basic Operation
4-28
RF-20A/20Axs
4.2.4 Setting the Output Ranges
Before setting the output ranges, set analog output connectors 1 and 2 in accordance with the equipment they
will be connected to.
^ "4.2.3 Setting the Analog Output Connectors" P.4-25
NOTE
Note that the output range settings only take effect if the analog output connector has been set for
RECORDER.
^ "4.2.3 Setting the Analog Output Connectors" P.4-25
When a Chromatopac Is Used as the Recorder:
Normally it is connected to the analog output connector set for INTEGRATOR.
^ "4.2.3 Setting the Analog Output Connectors" P.4-25
The range when a Chromatopac is used is normally set with the [ATTEN] setting at the Chromatopac, but
since the dynamic range at the instrument side is extremely wide, set gain and sensitivity at the instrument.
^ "4.2.7 Setting Gain" P.4-36
"4.2.8 Setting Sensitivity" P.4-38
Regardless of the range setting, the analog output connector set for INTEGRATOR outputs a voltage of 1 V
at the maximum value (1000) for fluorescent intensity.
When a Recorder Is Used:
Connect the recorder to the analog output connector set for RECORDER.
^ "4.2.3 Setting the Analog Output Connectors" P.4-25