4. Basic Operation
4-38
RF-20A/20Axs
4.2.8 Setting Sensitivity
This is the procedure for setting sensitivity. By combining the setting of sensitivity and gain, you can set the
measuring range.
^ "4.2.7 Setting Gain" P.4-36
NOTE
When using a Chromatopac, select the sensitivity of this instrument such that the noise becomes several
percent of the plot range when the [ATTEN] setting of the Chromatopac is [0]. Next, select an [ATTEN] setting
that ensures that the targeted peak is within the plot range. If the peak height is beyond the plot range even
when [ATTEN] of the Chromatopac is set to the lowest sensitivity, change the sensitivity of this instrument to
low sensitivity.
The settings for sensitivity are shown below.
On shipping from the factory, the setting is [2] (MED).
1
Press .
The initial screen will be displayed.
2
Press .
[PARAMETER] will be displayed.
3
Press .
[ch1] will be displayed.
Set Value Sensitivity Sensitivity Magnifications
1 HIGH Approx. × 1
2 MED Approx. × 32
3 LOW Approx. × 1024
EX200n m EM3 00nm
X e:c h 1 1000.00
PARAMETER
Enter to Select
ch1 EX350 EM450
Input 0,200-900