Owner manual
19
Troubleshooting
The following are general checkpoints; please call your local Shimpo representative or contact Shimpo Instruments directly for
further assistance.
The DFS does not turn on:
• Check all electrical components (power source … battery or power cord)
Error Codes are displayed:
• If any error codes appear on the display, turn power OFF and then ON
• If error code remains, see table below:
Calibration (continued)
DISPLAY CONDITION ACTION
EE PROM reading error Press RESET; unit will default to
factory settings
Memory data error Press RESET to clear memory
Calibration error Press RESET and repeat calibration
procedure
Memory unit error; occurs if the Change measurement units or clear
measurement unit (lb, Kg, N) is memory
changed while entering data into
memory
Overload condition –possible Remove excessive load; if the display
load cell damage does not return to normal operation,
send unit in for repair
MODEL SELECT NUMBER DISPLAY CALIBRATION WEIGHT
DFS-0.5 (R) 00 0.2 200g
DFS-1 (R) 01 0.5 500g
DFS-2 (R) 02 1.0 1000g
DFS-5 (R) 03 2.0 2.0kg
DFS-10 (R) 04 5.0 5.0kg
DFS-20 (R) 05 10.0 10.0kg
DFS-50 (R) 06 20.0 20.0kg
DFS-100 (R) 07 50.0 50.0kg
Please call Shimpo for information on additional sensors.