5 Specifications
Model Number DT-329
Units of measure FPM (flashes per minute) or Hz (frequency)
Measuring range 1 - 2,000Hz
60 - 120,000 FPM
Accuracy +/-0.02% (at 73°F [2C])
Resolution
of flash rate
setting in
internal
mode
Unit;
Hz
Measuring
range:
1
-
200 Hz
0.01Hz
Measuring
range:
1-2000 Hz
0.1Hz
Unit;
FPM
Measuring
range:
60 - 12,000 FPM
60.0 - 3,000.0
3,000.2 - 6,000.0
6,000.5 - 12,000.0
0.1FPM
0.2FPM
0.5FPM
Measuring
range:
60 - 120,000
FPM
60 - 30,000
30,002 - 60,000
60,005 - 100,000
100,010
-
120,000
1FPM
2FPM
5FPM
10FPM
Flashing time (Duty)
0.1° - 1.0°/36 setting can be changed by 0. inc rements
Mode Internal Set and change
flashing rate
Change flash rate by using dial, or “×2 and “1/2 key
Phase Shift Forward/backward by 3° increments
Unit change FPM or Hz
External
trigger
Flash rate
(Input frequency)
1 166 Hz
60 10,000 FPM
Input pulse width More than 50 us
Unit change FPM or Hz
Parameter
setting
Measuring range
1 - 2000Hz (60 - 120,000 FPM) range or
1 - 200Hz (60 - 12,000 FPM)
Trigger Edge Positive or Negative edge
Delay time/angle Delayed time : 0 - 999ms *1
Delayed angle : 0 - 359° *1
Input/ Output (I/F)
External Trigger Input
Photo Coupler Input, Current 12mA (approx.)
External Trigger output Photo Coupler (Open collector) Output
Voltage: DC30V or less, Current: 30mA or less.
Pulse width: approx.200us
Sensor Power Supply DC12V/max.40mA
*1
Built-in delay time = 60us