User's Manual
Table Of Contents
- SITRANS LR250 (FOUNDATION FIELDBUS)
- Legal information
- Table of contents
- 1 Introduction
- 2 Safety notes
- 3 Description
- 4 Installing/mounting
- 5 Connecting
- 6 Commissioning
- 7 Remote operation
- 7.1 Operating via AMS Device Manager
- 7.1.1 Functions in AMS Device Manager
- 7.1.2 Key features of AMS Device Manager Rev. 9.0
- 7.1.3 Adding a new device
- 7.1.4 Master Reset
- 7.1.5 Scan Device
- 7.1.6 Sensor calibration
- 7.1.7 Configuring a new device
- 7.1.8 Changing parameter settings using AMS Device Manager
- 7.1.9 Configure/Setup (Level Transducer Block-LTB)
- 7.1.10 Configure/Setup (Liquid Crystal Display Block-LCD)
- 7.1.11 Configure/Setup (Diagnostic Transducer Block-DIAG)
- 7.1.12 Configure/Setup (Resource Block - RESOURCE)
- 7.1.13 Device Diagnostics (Level Transducer Block - LTB)
- 7.1.14 Device Diagnostics (Liquid Crystal Display Block - LCD)
- 7.1.15 Device Diagnostics (Diagnostic Transducer Block - DIAG)
- 7.1.16 Device Diagnostics (Resource Block - RESOURCE)
- 7.1.17 Process Variables (Level Transducer Block - LTB)
- 7.1.18 Password Protection
- 7.1.19 AMS menu structure
- 7.1 Operating via AMS Device Manager
- 8 Parameter reference
- 9 Service and maintenance
- 10 Diagnosing and troubleshooting
- 11 Technical data
- 12 Dimension drawings
- 12.1 Threaded horn antenna
- 12.2 Threaded horn antenna with extension
- 12.3 Flanged horn antenna
- 12.4 Flanged horn antenna with extension
- 12.5 Flanged encapsulated antenna (2"/DN50/50A sizes only)
- 12.6 Flanged encapsulated antenna (3"/DN80/80A sizes and larger)
- 12.7 Threaded PVDF antenna
- 12.8 Threaded connection markings
- 12.9 Raised-Face flange per EN 1092-1 for flanged horn antenna
- 12.10 Raised-Face flange per EN 1092-1 for flanged encapsulated antenna
- 12.11 Flat-Face flange
- 12.12 Process connection tag (pressure rated versions)
- A Appendix A: Technical reference
- B Appendix B: Communications via Foundation Fieldbus
- C Appendix C: Certificates and support
- 13 List of abbreviations
- 14 LCD menu structure
- Glossary
- Index
![](/manual/siemens-canada-ltd-siemens-milltronics-process-instruments/lr250/2-1-user-s-manual-english/images/img-54.png)
Commissioning
6.1 Operating via the handheld programmer
SITRANS LR250 (FOUNDATION FIELDBUS)
Operating Instructions, 01/2014, A5E32221411-AB
53
6.1.3.3
Requesting an Echo Profile
•
In
PROGRAM
mode, navigate to:
Level Meter > Diagnostics
(3.) > Echo Profile (3.1
.)
•
Press
RIGHT arrow
to request a profile.
①
Confidence
⑦
Zoom
②
Distance from Low Calibration Point to vertical cross
-hair
⑧
Measure
③
Algorithm: tF
(trueFirst)
⑨
Ex
it
④
Distance from flange face to target
⑩
Exit icon selected
⑤
Pan left/right
- selected
⑪
Exit icon deselected
⑥
Pan up/down
● Use
UP
or
DOWN arrow
to scroll to an icon. When an icon is highlighted, that feature
becomes active.
● To move a cross-hair, press
RIGHT arrow
to increase the value,
LEFT arrow
to
decrease.
● To Zoom into an area, position the intersection of the cross-hairs at the center of that area,
select
Zoom
, and press
RIGHT arrow.
Press
LEFT arrow
to Zoom out.
● To update the profile, select
Measure
and press
RIGHT arrow.
● To return to the previous menu, select
Exit
then press
RIGHT arrow.