User's Manual

Table Of Contents
Commissioning
6.2 Application examples
SITRANS LR250 (FOUNDATION FIELDBUS)
Operating Instructions, 01/2014, A5E32221411-AB
59
Parameter type
Parameter Name
Options/
Values
Function
Independent
parameters
Vessel Shape (2.4.1.1.)
PARABOLIC
ENDS
Defines vessel shape.
PV (volume/level) Units
(2.3.3.)
L
liters
Level Unit (2.3.2.)
M
meters
Maximum Volume (2.4.1.2.)
8000
8000 liters
Dimension A (2.4.1.3.)
0.8
0.8 m (2.62 ft)
Dimension L (2.4.1.4.)
6
6 m (19.68 ft)
Loss of Echo (LOE) Timer
(2.3.6.)
120
120 seconds
Return to
Measurement:
press
Mode
to start normal operation.