User's Manual

Table Of Contents
Parameter reference
SITRANS LR250 (FOUNDATION FIELDBUS)
Operating Instructions, 01/2014, A5E32221411-AB
151
After Auto False Echo Suppression
false echo
material level
learned TVT
default TVT
Auto False Echo Suppression Range
echo marker
To use Auto False Echo Suppression via AMS Device Manager
note value calculated in step 1
and see
Auto False Echo Suppression (Page 85).
Auto False Echo Suppression Range (2.5.10.2.)
Defines the endpoint of the Learned TVT distance. Units are defined in
Unit (2.3.1.)
.
Values
Range: 0.00 to 30.00 m
Default: 1.00 m
1. Press
RIGHT arrow
to open Edit mode.
2. Enter the new value and press
RIGHT arrow
to accept it.
3. Set
Auto False Echo Suppression (2.5.10.1.)
.